Separation Methods: Extraction, Chromatography and Ion Exchange

CSIR NET chemistry: solvent extraction and D vs pH, plate theory, van Deemter, resolution, GC, HPLC, ion exchange, size exclusion and electrophoresis.

  • 24 min read
  • Updated
On this page · 17 sections
  1. Prerequisites
  2. The Hook: Why Three Small Washes Beat One Big One
  3. 1. Solvent Extraction
  4. 2. Chromatography: Common Language
  5. 3. Planar Chromatography: Paper and TLC
  6. 4. Gas Chromatography (GC)
  7. 5. Liquid Chromatography (HPLC)
  8. 6. Ion Exchange
  9. 7. Electrophoresis in Brief
  10. 8. Other Separations Worth Knowing
  11. Memory Tricks
  12. Don’t Confuse
  13. Common Mistakes
  14. Practice
  15. Quick Revision Table
  16. Key Takeaways
  17. What’s Next

Prerequisites

The Hook: Why Three Small Washes Beat One Big One

1. Solvent Extraction

Partition coefficient and distribution ratio

When a solute S distributes between an aqueous phase and an immiscible organic phase, the partition (distribution) coefficient refers to one chemical species:

$$K_D = \frac{[\text{S}]_{\text{org}}}{[\text{S}]_{\text{aq}}}$$

The distribution ratio D refers to the total analytical concentration, whatever form the solute is in:

$$D = \frac{C_{\text{org}}}{C_{\text{aq}}} = \frac{\text{all forms of S in the organic phase}}{\text{all forms of S in the aqueous phase}}$$

\( K_D \) is a constant at fixed temperature (Nernst distribution law). D changes with pH, with complexing agents and with ionic strength, because those change how much of the solute is in the extractable form. Experiments measure D.

Fraction remaining after repeated extraction

For an aqueous volume \( V_{\text{aq}} \) extracted n times with fresh portions of organic solvent of volume \( V_{\text{org}} \) each, the fraction q left in water is

$$\boxed{q_n = \left(\frac{V_{\text{aq}}}{V_{\text{aq}} + D\,V_{\text{org}}}\right)^n}$$

and the percentage extracted after a single extraction is

$$\%E = \frac{100\,D}{D + V_{\text{aq}}/V_{\text{org}}}$$

With equal volumes, \( \%E = 100D/(D+1) \): D = 1 gives 50%, D = 10 gives 90.9%, D = 100 gives 99.0%.

Extracting a weak acid: D depends on pH

Only neutral HA enters a non-polar solvent; A⁻ stays in water. With \( K_D \) for HA and acid constant \( K_a \):

$$\boxed{D = \frac{K_D\,[\text{H}^+]}{[\text{H}^+] + K_a}}$$
  • At pH well below pKa, \( D \to K_D \) (all HA, all extractable).
  • At pH = pKa, \( D = K_D/2 \).
  • At pH well above pKa, \( D \approx K_D[\text{H}^+]/K_a \): it falls tenfold per pH unit.

For \( K_D = 20 \) and \( K_a = 1.0 \times 10^{-5} \): D = 19.8 at pH 3, 10.0 at pH 5, 0.198 at pH 7. This is how a carboxylic acid is pulled out of an organic layer into aqueous NaHCO₃ and then back after acidifying. A weak base B behaves the other way round: it extracts at high pH, where it is unprotonated.

Metal chelate extraction

A metal ion is charged and hydrated, so it does not enter an organic solvent until it is converted into a neutral, hydrophobic species. The two standard routes are:

RouteExtracted speciesExamples
Chelate formationneutral ML_n with a weak-acid chelating ligand HL8-hydroxyquinoline (oxine), dithizone, acetylacetone, cupferron, dimethylglyoxime (Ni)
Ion associationneutral ion pair or solvated saltFe(III) from concentrated HCl into ether as H⁺(solvated)·FeCl₄⁻; large cations such as tetraphenylarsonium with large anions

For \( \text{M}^{n+}(\text{aq}) + n\text{HL(org)} \rightleftharpoons \text{ML}_n(\text{org}) + n\text{H}^+(\text{aq}) \), with extraction constant \( K_{\text{ex}} \) and the chelating agent in excess:

$$\boxed{D = K_{\text{ex}}\,\frac{[\text{HL}]_{\text{org}}^{\,n}}{[\text{H}^+]^{n}}} \qquad \log D = \log K_{\text{ex}} + n\log[\text{HL}]_{\text{org}} + n\,\text{pH}$$

Three consequences that Part C tests:

  1. A plot of log D against pH is a straight line of slope n, the charge of the metal ion. Slope 2 means M²⁺, slope 3 means M³⁺.
  2. The pH₁/₂, where D = 1 (50% extracted with equal volumes), is \( \text{pH}_{1/2} = -\frac{1}{n}\log K_{\text{ex}} - \log[\text{HL}]_{\text{org}} \). A tenfold increase in reagent concentration lowers pH₁/₂ by one unit.
  3. Two metals with different \( K_{\text{ex}} \) have different pH₁/₂, so they can be separated by buffering between them. Masking agents (EDTA, cyanide, tartrate) form water-soluble complexes and keep an interfering metal in the aqueous phase.

2. Chromatography: Common Language

All chromatography distributes solutes between a stationary phase and a mobile phase that moves past it. A solute that spends more time in the stationary phase moves more slowly.

graph TD
    A[Chromatography] --> B[Mobile phase: gas]
    A --> C[Mobile phase: liquid]
    A --> X[Mobile phase: supercritical fluid]
    B --> B1[GC: gas-liquid partition
gas-solid adsorption] C --> C1[Planar: paper, TLC] C --> C2[Column: HPLC] C2 --> D1[Adsorption / normal phase] C2 --> D2[Partition / reversed phase] C2 --> D3[Ion exchange] C2 --> D4[Size exclusion] C2 --> D5[Affinity]

Retention parameters

QuantityDefinitionMeaning
Retention time \( t_R \)time from injection to peak maximumidentity, under fixed conditions
Dead time \( t_M \)time for an unretained speciestime spent in the mobile phase
Adjusted retention time \( t_R' \)\( t_R - t_M \)time spent in the stationary phase
Retention (capacity) factor\( k = \dfrac{t_R - t_M}{t_M} \)moles in stationary / moles in mobile phase; ideally 1 to 10
Selectivity factor\( \alpha = \dfrac{k_B}{k_A} = \dfrac{t'_{R,B}}{t'_{R,A}} \) (B the later peak, so α ≥ 1)how differently the phases treat A and B
Retardation factor (planar)\( R_f = \dfrac{\text{distance moved by spot}}{\text{distance moved by solvent front}} \)always between 0 and 1

The retention factor is related to the distribution constant: \( k = K\,V_S/V_M \), where \( V_S \) and \( V_M \) are the volumes of stationary and mobile phase.

Efficiency: plate number and plate height

Plate theory treats the column as N imaginary equilibrium stages. N is measured from the peak:

$$\boxed{N = 16\left(\frac{t_R}{W}\right)^2 = 5.54\left(\frac{t_R}{W_{1/2}}\right)^2} \qquad \boxed{H = \frac{L}{N}}$$

W is the baseline width (from tangents to the inflection points, equal to 4σ for a Gaussian peak), \( W_{1/2} \) the width at half height, L the column length and H the height equivalent to a theoretical plate (HETP). Small H means an efficient column. t_R and W must be in the same unit.

Why peaks broaden: the van Deemter equation

Rate theory explains band broadening as the sum of three contributions, each depending differently on the linear flow velocity u of the mobile phase:

$$\boxed{H = A + \frac{B}{u} + C\,u}$$
TermNamePhysical causeHow to reduce it
Aeddy (multipath) diffusionmolecules take paths of different length through the packing; independent of usmall, uniform particles; open tubular (capillary) columns have no A term
B/ulongitudinal diffusionsolute diffuses along the column axis while waiting; worse at low flowfaster flow; less important in liquids, where diffusion coefficients are about 10⁴ times smaller than in gases
Curesistance to mass transferequilibration between phases is not instantaneous; worse at high flowthin stationary film, small particles, low viscosity, higher temperature

Setting \( dH/du = 0 \) gives the optimum velocity and the minimum plate height:

$$\boxed{u_{\text{opt}} = \sqrt{B/C}} \qquad \boxed{H_{\min} = A + 2\sqrt{BC}}$$

The curve of H against u is a hyperbola-like plot with a minimum. To the left (slow flow) the B term dominates; to the right (fast flow) the C term dominates, and the curve rises almost linearly with slope C.

Resolution

Resolution measures how well two adjacent peaks are separated:

$$\boxed{R_s = \frac{2\,(t_{R,B} - t_{R,A})}{W_A + W_B}}$$

\( R_s = 1.5 \) is taken as baseline resolution; at \( R_s = 1.0 \) the peaks still overlap slightly. The Purnell equation splits resolution into the three things you can control:

$$\boxed{R_s = \frac{\sqrt{N}}{4}\left(\frac{\alpha - 1}{\alpha}\right)\left(\frac{k_B}{1 + k_B}\right)}$$
FactorHow to change itEffect
Efficiency \( \sqrt{N} \)longer column, smaller particles, optimum flow\( R_s \propto \sqrt{L} \): doubling resolution needs four times the length (and about four times the time)
Selectivity αchange stationary phase, mobile-phase composition, pH, temperaturemost powerful: small change in α gives a big change in R_s
Retention \( k_B \)solvent strength (HPLC) or temperature (GC)helps up to k of about 5 to 10, then little further gain and much longer runs

The general elution problem

A sample with components spanning a wide range of k gives early peaks crowded together and late peaks broad and flat. The fix is to change conditions during the run:

  • GC: temperature programming (raise column temperature during the run).
  • HPLC: gradient elution (increase the strength of the mobile phase during the run).

3. Planar Chromatography: Paper and TLC

  • Paper chromatography is partition chromatography: the stationary phase is water held by cellulose, the mobile phase an organic solvent mixture.
  • Thin-layer chromatography (TLC) on silica gel or alumina is mainly adsorption chromatography. Polar compounds adsorb strongly on silica and have low R_f; a more polar eluent raises all R_f values.
  • Two-dimensional development (second solvent at right angles) resolves mixtures such as amino acids that one solvent cannot.
  • Colourless spots are located with UV (fluorescent indicator in the layer), iodine vapour or a spray reagent (ninhydrin for amino acids).

R_f depends on the layer, the solvent, temperature and chamber saturation, so identification needs a standard run on the same plate.

4. Gas Chromatography (GC)

Mobile phase: an inert carrier gas (He, N₂, H₂). Stationary phase: a high-boiling liquid (for example a polysiloxane) coated on the inner wall of a fused-silica capillary or on a packed support. The sample must be volatile and thermally stable, or made so by derivatisation (silylation, methylation).

Detectors

DetectorPrincipleResponds toDoes not respond (or responds weakly) to
TCD (thermal conductivity)analyte changes thermal conductivity of carrier gasuniversal; non-destructivelow sensitivity
FID (flame ionisation)organic carbon burnt in H₂/air flame gives ions; current measuredorganic compounds with C–H bonds; very sensitive, wide linear rangeH₂O, CO₂, CO, N₂, O₂, noble gases; carbonyl and halogenated carbons respond less
ECD (electron capture)β-source (⁶³Ni) ionises carrier; electronegative analyte captures electrons, current dropshalogenated compounds, nitro compounds, conjugated carbonyls (pesticides, PCBs)hydrocarbons, alcohols
NPD / thermionicalkali-salt beadN- and P-containing compoundshydrocarbons
FPD (flame photometric)emission from S and P species in a flameS and P compoundsmost others
MSmass spectrometeruniversal with structural identity—

Retention index

Retention times change between instruments, so GC identities are reported as Kovats retention indices, which place a compound on a scale defined by n-alkanes (each n-alkane has I = 100 × carbon number). Under isothermal conditions, for a compound eluting between alkanes with n and n + 1 carbons:

$$I = 100\left[n + \frac{\log t'_{R,x} - \log t'_{R,n}}{\log t'_{R,n+1} - \log t'_{R,n}}\right]$$

Example: adjusted retention times n-octane 4.00 min, n-nonane 8.00 min, unknown 5.66 min. \( I = 100[8 + \log(5.66/4.00)/\log 2] = 100(8 + 0.500) = 850 \).

5. Liquid Chromatography (HPLC)

ModeStationary phaseMobile phaseElution orderTypical use
Normal phasepolar (silica, alumina, bonded –CN, –NH₂)non-polar (hexane with a little polar modifier)least polar firstisomers, lipids
Reversed phasenon-polar (C18 or C8 bonded silica)polar (water with methanol or acetonitrile)most polar firstmost organic and pharmaceutical analysis
Ion exchangecharged resinaqueous bufferweakly bound ions firstinorganic ions, amino acids
Size exclusion (gel permeation / gel filtration)porous gelany solvent that dissolves the samplelargest firstpolymer molar mass distribution, proteins
Affinityimmobilised specific ligandbuffer, then eluting agentnon-binders firstenzymes, antibodies

In reversed phase, adding more organic solvent makes the mobile phase stronger: retention of every component falls. In normal phase, adding a more polar modifier does the same. The order of eluting strength on silica (the eluotropic series) runs roughly pentane < CCl₄ < toluene < CH₂Cl₂ < diethyl ether < ethyl acetate ≈ acetone ≈ acetonitrile < methanol < water.

Size exclusion is the only mode where retention does not depend on chemical interaction. Molecules larger than the largest pores are fully excluded and elute together at the void volume; molecules small enough to enter every pore elute together at the total permeation volume; in between, elution volume falls roughly linearly with log(molar mass), which is how a column is calibrated with standards.

HPLC detectors: UV–visible (most common, needs a chromophore), fluorescence (very sensitive, selective), refractive index (universal but insensitive and unusable with gradients), conductivity (ions), electrochemical, and mass spectrometric.

6. Ion Exchange

Resins

TypeFunctional groupExchangesWorks over
Strong-acid cation–SO₃⁻H⁺cationsalmost all pH
Weak-acid cation–COO⁻H⁺cationsonly where –COOH is ionised (higher pH)
Strong-base anion–CH₂N(CH₃)₃⁺OH⁻ (quaternary ammonium)anionsalmost all pH
Weak-base anion–NH₂, –NHR (protonated)anionsonly where the amine is protonated (lower pH)

Most are cross-linked polystyrene–divinylbenzene. Capacity is quoted in milliequivalents per gram of dry resin.

$$\text{R–SO}_3^-\text{H}^+ + \text{Na}^+ \rightleftharpoons \text{R–SO}_3^-\text{Na}^+ + \text{H}^+$$

Selectivity rules

  1. Higher charge binds more strongly: \( \text{Th}^{4+} > \text{La}^{3+} > \text{Ca}^{2+} > \text{Na}^+ \).
  2. Same charge: the ion with the smaller hydrated radius binds more strongly. For the alkali metals the bare-ion order is reversed by hydration: \( \text{Cs}^+ > \text{Rb}^+ > \text{K}^+ > \text{Na}^+ > \text{Li}^+ \). Likewise \( \text{Ba}^{2+} > \text{Sr}^{2+} > \text{Ca}^{2+} > \text{Mg}^{2+} \).
  3. For anions on a strong-base resin, large polarisable, weakly hydrated anions bind more strongly: \( \text{I}^- > \text{Br}^- > \text{Cl}^- > \text{F}^- \).
  4. Dilute solutions favour the more highly charged ion (electroselectivity); concentrated solutions reduce this preference, which is why concentrated acid or brine regenerates a resin.

Classic separations

  • Lanthanides on a cation exchanger eluted with a complexing agent (citrate, EDTA, α-hydroxyisobutyrate): the ions differ little in their affinity for the resin, but complex strength with the eluent increases from La to Lu as the ionic radius contracts. The heaviest lanthanide forms the strongest complex and leaves the resin first: Lu elutes first, La last. (See the lanthanide contraction in your inner-transition chemistry.)
  • Transition metals on an anion exchanger from concentrated HCl: metals that form anionic chloro-complexes (CoCl₄²⁻, FeCl₄⁻, ZnCl₄²⁻) are held; Ni²⁺, which does not form a stable anionic chloro-complex, passes straight through. Lowering the HCl concentration step by step then releases Co, then Fe, with Zn held longest.
  • Deionisation: water through H⁺-form cation and OH⁻-form anion resins (or a mixed bed) loses all ions as H⁺ + OH⁻ → H₂O.
  • Ion chromatography: anions separated on a low-capacity anion column with a carbonate or hydroxide eluent, followed by a suppressor that converts the eluent to weakly conducting H₂CO₃ or water, so that the analyte ions can be detected by conductivity.

7. Electrophoresis in Brief

Charged species migrate in an electric field with velocity \( v = \mu E \), where μ is the electrophoretic mobility (roughly charge divided by size). In capillary zone electrophoresis a fused-silica capillary has ionised silanol groups; the cation layer at the wall drags the bulk solution towards the cathode (electro-osmotic flow, EOF), usually faster than any ion’s own migration. All species therefore reach a detector at the cathode end, in the order:

cations (fastest first, small and highly charged) → neutrals (all together, at the EOF velocity) → anions (largest mobility last).

Neutral molecules are not separated in plain CZE; adding a surfactant above its CMC (micellar electrokinetic chromatography) separates them by partition into the micelles.

8. Other Separations Worth Knowing

MethodBasisExample
Precipitation with organic reagentsselective insoluble chelatesdimethylglyoxime gives red Ni(dmg)₂ in ammoniacal solution; oxine precipitates Al³⁺ and Mg²⁺ at different pH
Maskingkeeps an ion in solution as a soluble complexcyanide masks Cu, Zn, Ni in EDTA titrations; fluoride masks Fe³⁺ and Al³⁺
Distillation / volatilisationvolatile speciesNH₃ in Kjeldahl analysis; As as AsCl₃; Os and Ru as tetroxides
Solid-phase extractionsmall sorbent cartridgespre-concentration of trace organics from water

Memory Tricks

Don’t Confuse

ThisvsThat
Partition coefficient \( K_D \): one species, constantdistribution ratio D: all forms, changes with pH and ligands
Retention factor \( k = (t_R - t_M)/t_M \)retardation factor \( R_f \) (planar, distance ratio, 0 to 1)
Selectivity α (ratio of k values)resolution R_s (includes peak widths)
N from baseline width: factor 16N from half-height width: factor 5.54
Normal phase: polar stationary, polar elutes lastreversed phase: non-polar stationary, polar elutes first
FID: blind to water and CO₂TCD: universal, less sensitive
ECD: halogenated, nitro compoundsFPD: sulfur and phosphorus
Cation exchanger: smaller hydrated ion binds more (Cs⁺ > Li⁺)gas-phase intuition that the smaller bare ion should bind more
Size exclusion: no chemical interaction, largest firstadsorption: retention by surface binding

Common Mistakes

Practice

All questions below are original practice questions written for this page. They are not from any CSIR NET paper.

Level 1: Foundation

  1. State the difference between the partition coefficient and the distribution ratio.
  2. In size-exclusion chromatography, which elutes first: a protein of 150 kDa or one of 15 kDa?
  3. Write the van Deemter equation and say which term is absent in an open tubular column.
Answers
  1. \( K_D \) is the concentration ratio of a single species between the phases; D is the ratio of total analytical concentrations, summed over all forms (ionised, complexed), and changes with pH and ligand concentration.
  2. The 150 kDa protein. Larger molecules enter fewer pores and travel the shorter path.
  3. \( H = A + B/u + Cu \). The A (eddy diffusion, multiple path) term is absent because there is no packing.

Level 2: Part B Style

Practice Q1. A solute has D = 5 between ether and water. 100 mL of its aqueous solution is extracted three times with 20 mL portions of ether. The percentage extracted is:

(a) 75.0 (b) 83.3 (c) 87.5 (d) 96.3

Answer: (c) 87.5

\( q_3 = [100/(100 + 5 \times 20)]^3 = (0.5)^3 = 0.125 \), so 87.5% is extracted. Option (a) is the result of a single 60 mL extraction: \( 100/(100 + 300) = 0.25 \) remaining. Option (b) is a single extraction with 100 mL of ether, \( 100D/(D+1) = 83.3\% \).

Practice Q2. A peak has a retention time of 5.0 min and a baseline width of 0.25 min on a 25 cm column. The plate height is closest to:

(a) 3.9 μm (b) 39 μm (c) 0.39 mm (d) 64 μm

Answer: (b) 39 μm

\( N = 16(5.0/0.25)^2 = 6400 \). \( H = L/N = 25\ \text{cm}/6400 = 3.9 \times 10^{-3}\ \text{cm} = 39\ \mu\text{m} \). Options (a) and (c) are unit slips by a factor of ten.

Practice Q3. A mixture of La³⁺, Eu³⁺ and Lu³⁺ is loaded on a strong-acid cation-exchange column and eluted with ammonium α-hydroxyisobutyrate. The order of elution is:

(a) La, Eu, Lu (b) Lu, Eu, La (c) Eu, La, Lu (d) all three together, since they have the same charge

Answer: (b) Lu, Eu, La

Complex stability with the eluent rises across the series as the ionic radius falls (lanthanide contraction). The most strongly complexed ion, Lu³⁺, spends the most time in the mobile phase and elutes first; La³⁺ elutes last. Option (d) ignores the role of the complexing eluent.

Practice Q4. The most suitable GC detector for trace chlorinated pesticides in a hydrocarbon-rich soil extract is:

(a) thermal conductivity (b) flame ionisation (c) electron capture (d) refractive index

Answer: (c) electron capture

The ECD is highly sensitive to electronegative, halogen-containing compounds and almost blind to hydrocarbons, so the matrix does not swamp the signal. FID would respond strongly to the hydrocarbons; TCD is too insensitive; RI is an HPLC detector.

Practice Q5. On a C18 reversed-phase column with methanol–water mobile phase, which change shortens the retention time of a neutral aromatic analyte?

(a) increasing the water content (b) increasing the methanol content (c) changing to a silica column with the same mobile phase (d) lowering the flow rate

Answer: (b)

More organic solvent makes the mobile phase more like the non-polar stationary phase, so the analyte partitions more into the mobile phase and k falls. (a) does the opposite; (d) lengthens all retention times without changing k.

Level 3: Part C Style

Practice Q6. A metal ion Mⁿ⁺ is extracted into chloroform by a chelating agent HL held at 0.010 M in the organic phase (equal phase volumes). The measured distribution ratios are log D = −1.0 at pH 2.0 and log D = +1.0 at pH 3.0. Which statements are correct?

  1. The metal ion is divalent.
  2. log K_ex = −1.0.
  3. pH₁/₂ is 2.5 at this reagent concentration.
  4. Raising [HL]_org to 0.10 M moves pH₁/₂ to 3.5.

(a) 1, 2 and 3 only (b) 1 and 4 only (c) 2, 3 and 4 only (d) 1, 2, 3 and 4

Answer: (a) 1, 2 and 3 only

\( \log D = \log K_{ex} + n\log[\text{HL}]_{org} + n\,\text{pH} \).

  • Slope: \( (1.0 - (-1.0))/(3.0 - 2.0) = 2 \), so n = 2: statement 1 true.
  • At pH 2.0: \( -1.0 = \log K_{ex} + 2(-2) + 2(2.0) \), so \( \log K_{ex} = -1.0 \): statement 2 true.
  • pH₁/₂ (log D = 0): \( 0 = -1.0 - 4 + 2\,\text{pH} \), pH = 2.5: statement 3 true.
  • With [HL] = 0.10 M, log[HL] = −1: \( 0 = -1.0 - 2 + 2\,\text{pH} \), pH₁/₂ = 1.5. More reagent lowers pH₁/₂ by one unit: statement 4 false.

Practice Q7. For a packed GC column the van Deemter constants are A = 0.08 mm, B = 0.60 mm cm s⁻¹ and C = 0.15 mm s cm⁻¹. Which pair gives the optimum linear velocity and the minimum plate height?

(a) 2.0 cm s⁻¹, 0.68 mm (b) 4.0 cm s⁻¹, 0.38 mm (c) 2.0 cm s⁻¹, 0.38 mm (d) 0.5 cm s⁻¹, 0.68 mm

Answer: (a)

\( u_{opt} = \sqrt{B/C} = \sqrt{0.60/0.15} = \sqrt{4} = 2.0 \) cm s⁻¹.

\( H_{min} = A + 2\sqrt{BC} = 0.08 + 2\sqrt{0.09} = 0.08 + 0.60 = 0.68 \) mm.

Check by substitution: \( H(2.0) = 0.08 + 0.60/2.0 + 0.15 \times 2.0 = 0.08 + 0.30 + 0.30 = 0.68 \) mm. At the optimum the B and C terms are equal. Option (b) takes u = B/C, forgetting the square root (at 4.0 cm s⁻¹ the true H would be 0.83 mm); option (c) has the right velocity but drops one of the two equal terms (0.08 + 0.30); option (d) inverts the ratio, \( \sqrt{C/B} = 0.5 \).

Practice Q8. Two compounds give R_s = 1.0 on a 15 cm HPLC column. Assume H, α and k do not change when the column is lengthened. Which statements are correct?

  1. A 30 cm column gives R_s = 2.0.
  2. About 34 cm of column is needed for R_s = 1.5.
  3. On the longer column the analysis takes proportionally longer.
  4. Changing the mobile phase to raise α from 1.05 to 1.10 nearly doubles R_s at constant N and k.

(a) 1 and 3 only (b) 2, 3 and 4 only (c) 2 and 4 only (d) 1, 2, 3 and 4

Answer: (b) 2, 3 and 4 only
  • \( R_s \propto \sqrt{N} \propto \sqrt{L} \). Doubling L gives \( \sqrt{2} \times 1.0 = 1.41 \), not 2.0: statement 1 false.
  • For R_s = 1.5: \( L = 15 \times (1.5/1.0)^2 = 15 \times 2.25 = 33.75 \) cm, about 34 cm: statement 2 true.
  • Retention time scales with L at the same flow velocity: statement 3 true.
  • The selectivity term \( (\alpha - 1)/\alpha \) goes from 0.05/1.05 = 0.0476 to 0.10/1.10 = 0.0909, a factor of 1.91: statement 4 true. This is why selectivity is the cheapest lever.

Quick Revision Table

IdeaFormula or fact
Distribution ratioD = C_org/C_aq (all forms); K_D for one species
Repeated extraction\( q_n = [V_{aq}/(V_{aq} + DV_{org})]^n \); %E = 100D/(D + V_aq/V_org)
Weak acidD = K_D[H⁺]/([H⁺] + K_a)
Metal chelatelog D = log K_ex + n log[HL] + n pH; slope n; pH₁/₂ falls 1 unit per tenfold HL
Retentionk = (t_R − t_M)/t_M; α = k_B/k_A; R_f = spot/front
EfficiencyN = 16(t_R/W)² = 5.54(t_R/W₁/₂)²; H = L/N
van DeemterH = A + B/u + Cu; u_opt = √(B/C); H_min = A + 2√(BC)
ResolutionR_s = 2Δt_R/(W_A + W_B); 1.5 = baseline; R_s ∝ √N, (α − 1)/α, k/(1 + k)
GC detectorsTCD universal; FID organics, blind to H₂O/CO₂; ECD halogens; FPD S, P; NPD N, P
Kovats indexn-alkane = 100 × C number; log interpolation of adjusted t_R
HPLC modesnormal: polar last; reversed: polar first; SEC: largest first
Ion exchangehigher charge binds more; same charge: Cs⁺ > K⁺ > Na⁺ > Li⁺; I⁻ > Br⁻ > Cl⁻ > F⁻
Lanthanidescomplexing eluent on cation exchanger: Lu first, La last
CZEEOF to cathode; cations, then neutrals, then anions

Key Takeaways