Spectroscopic Methods of Analysis

CSIR NET chemistry: Beer–Lambert law and deviations, multicomponent analysis, Job's method, AAS and its interferences, flame and ICP emission, fluorescence.

  • 21 min read
  • Updated
On this page · 15 sections
  1. Prerequisites
  2. The Hook: The Same Sodium Line, Two Opposite Instruments
  3. 1. The Beer–Lambert Law
  4. 2. Quantitative Spectrophotometry in Practice
  5. 3. Atomic Absorption Spectroscopy (AAS)
  6. 4. Atomic Emission: Flame Photometry and ICP
  7. 5. Molecular Fluorescence
  8. 6. Other Optical Methods in One Table
  9. Memory Tricks
  10. Don’t Confuse
  11. Common Mistakes
  12. Practice
  13. Quick Revision Table
  14. Key Takeaways
  15. What’s Next

Prerequisites

The Hook: The Same Sodium Line, Two Opposite Instruments

1. The Beer–Lambert Law

Monochromatic light of radiant power \( P_0 \) passes through a path b of a solution of concentration c and leaves with power P.

$$T = \frac{P}{P_0} \qquad \boxed{A = -\log_{10} T = \varepsilon\,b\,c}$$
SymbolNameUsual unit
Ttransmittance (×100 for %T)none
Aabsorbancenone
εmolar absorption coefficient (molar absorptivity)L mol⁻¹ cm⁻¹
bpath lengthcm
cconcentrationmol L⁻¹

Useful pairs: T = 10% ↔ A = 1.00; T = 1% ↔ A = 2.00; T = 50% ↔ A = 0.301. Absorbance is additive: for a mixture, \( A = b\sum_i \varepsilon_i c_i \) at each wavelength.

Deviations from Beer’s law

The law assumes monochromatic light, dilute solution and absorbers that do not change identity with concentration. When a calibration plot of A against c curves, the cause is one of three kinds:

KindCauseTypical sign
Realat high concentration (above about 0.01 M) solute–solute interactions and refractive-index changes alter εcurvature at high c only
Chemicalthe absorber takes part in an equilibrium that shifts with dilution: weak-acid dissociation, dimerisation, complex dissociation. Example: dichromate–chromate, \( \text{Cr}_2\text{O}_7^{2-} + \text{H}_2\text{O} \rightleftharpoons 2\text{HCrO}_4^- \), so an unbuffered solution does not obey Beer’s law at most wavelengthspositive or negative curvature; disappears at an isosbestic point or in a buffer
Instrumentalpolychromatic light (finite bandwidth) where ε varies across the band; stray light reaching the detector; mismatched cellsalways negative deviation at high A for stray light

Where the reading is most precise

If the dominant error is a constant uncertainty in reading T (typical of older instruments), the relative concentration error is

$$\frac{\Delta c}{c} = \frac{0.434\,\Delta T}{T\,\log T}$$

which is smallest at T = 1/e = 36.8%, i.e. A = 0.434. In practice absorbances between about 0.2 and 0.8 give the best precision; dilute or change the path length to land there.

Isosbestic points

If two absorbing species interconvert (HIn ⇌ In⁻, or free ligand ⇌ complex) and only these two are present, every spectrum in a series of different pH values or mixing ratios passes through the wavelength where their ε values are equal. This isosbestic point is evidence that only two species are involved; its loss signals a third species or a side reaction.

2. Quantitative Spectrophotometry in Practice

Calibration, sensitivity and detection limit

  • External calibration: measure standards, fit A = m c + intercept, read unknowns from the line.
  • Sensitivity is the slope m of the calibration line.
  • Limit of detection (common IUPAC-style convention): the concentration giving a signal three times the standard deviation of the blank, \( c_{\text{LOD}} = 3s_{\text{blank}}/m \). The limit of quantitation uses 10 s.

Example: blank standard deviation 0.002 absorbance units, slope 0.050 per ppm: LOD = 3 × 0.002/0.050 = 0.12 ppm.

Standard addition

When the sample matrix changes the signal (viscosity in a nebuliser, other absorbers, chemical interference), standards made in pure water give the wrong slope. Standard addition puts the standards into the sample itself, so the matrix is the same.

An internal standard is different: a known amount of a different element or compound is added to every sample and standard, and the ratio of signals is plotted. It corrects for fluctuations in injection volume, nebulisation rate or flame conditions (common in flame emission, ICP and chromatography).

Two-component mixtures

Because absorbances add, two components X and Y can be determined from measurements at two wavelengths, ideally the two absorption maxima:

$$A_{\lambda_1} = \varepsilon_{X,1}\,b\,c_X + \varepsilon_{Y,1}\,b\,c_Y \qquad A_{\lambda_2} = \varepsilon_{X,2}\,b\,c_X + \varepsilon_{Y,2}\,b\,c_Y$$

Two equations, two unknowns. The classic example is Cr(VI) and Mn(VII) (dichromate and permanganate) in steel analysis.

Stoichiometry of complexes

MethodProcedureResult
Job’s method of continuous variationequimolar M and L solutions mixed in varying ratios, total moles constant; plot A (corrected for M and L) against mole fraction of L, \( x_L \)maximum at \( x_L = n/(1+n) \) for ML_n: 0.5 for ML, 0.67 for ML₂, 0.75 for ML₃
Mole-ratio methodfixed [M], increasing [L]; plot A against [L]/[M]two straight lines meeting at the ratio n; a rounded corner means a weak complex
Slope-ratio methodtwo series, one with excess L, one with excess Mratio of slopes gives n; suits weak complexes

The curvature near the break or maximum also carries the formation constant: the sharper the peak, the more stable the complex.

Photometric titrations

Absorbance is followed during a titration at a wavelength where one species absorbs. Because A is linear in concentration, each branch is a straight line (after correcting for dilution) and the end point is their intersection.

Absorbing speciesShape of A versus titrant volume
Titrant onlyflat at zero, then rises after the end point
Product onlyrises to the end point, then flat
Analyte onlyfalls to the end point, then flat at zero
Analyte and titrant (product does not absorb)falls, then rises: V-shape

Photometric end points work even when the reaction is not complete near equivalence, because extrapolation uses points away from it. Example: Cu²⁺ titrated with EDTA at a wavelength where only the Cu–EDTA complex absorbs gives a rising line followed by a flat one.

3. Atomic Absorption Spectroscopy (AAS)

The sample solution is nebulised into a flame (or dried and atomised in a graphite furnace), turning the analyte into free ground-state atoms. Light at the element’s resonance line is passed through and absorbance is measured.

Why a hollow-cathode lamp

Atomic absorption lines are extremely narrow (of the order of thousandths of a nanometre). A continuum source passed through a monochromator would present a band far wider than the line, and the tiny fraction absorbed would barely change the signal. A hollow-cathode lamp whose cathode is made of the analyte element emits that element’s lines, even narrower than the absorption lines because the lamp runs at low pressure and temperature. So each element needs its own lamp (or a multi-element lamp).

The atomic line width comes from three sources: the natural (uncertainty) width, Doppler broadening from thermal motion (grows with temperature), and pressure (collisional) broadening.

Atomisers

AtomiserFeatures
Air–acetylene flamemost common; suitable for many metals
Nitrous oxide–acetylene flamehotter and more reducing; needed for elements that form refractory oxides (Al, Si, Ti, V, Mo)
Graphite furnace (electrothermal)the whole sample is atomised in a small tube in a programmed dry–ash–atomise sequence; much lower detection limits than a flame, microlitre samples, but slower and more matrix effects
Cold-vapourHg reduced to the element by SnCl₂ and swept as vapour
Hydride generationAs, Se, Sb, Bi, Sn converted to volatile hydrides by NaBH₄ and decomposed in a heated cell

Interferences and their cures

InterferenceWhat happensCure
Spectral (line overlap)another element absorbs at the same wavelengthchoose another line
Spectral (background)broad molecular absorption or scattering by particlesbackground correction: deuterium-lamp continuum, Zeeman effect, or Smith–Hieftje (lamp pulsing)
Chemicalanalyte forms a compound that does not atomise: Ca²⁺ with phosphate, sulfate, silicate or aluminate gives low resultsreleasing agent (La³⁺ or Sr²⁺ binds the phosphate preferentially); protective agent (EDTA, 8-hydroxyquinoline forms a volatile complex with the analyte); hotter N₂O–C₂H₂ flame
Ionisationin a hot flame, atoms of low ionisation energy (alkali metals) are partly ionised, so fewer neutral atoms absorbionisation suppressor: add a large excess of a more easily ionised element (K, Cs) to raise the electron concentration and push M ⇌ M⁺ + e⁻ back
Physical (matrix)viscosity, surface tension, dissolved solids change nebulisationmatrix-matched standards or standard addition

4. Atomic Emission: Flame Photometry and ICP

The Boltzmann distribution

At temperature T, the ratio of atoms in an excited state j to those in the ground state 0 is

$$\boxed{\frac{N_j}{N_0} = \frac{g_j}{g_0}\exp\left(-\frac{\Delta E}{kT}\right)}$$

where g are statistical weights (degeneracies). Emission intensity is proportional to \( N_j \); absorption to \( N_0 \).

Consequences:

  1. Emission needs tight temperature control; absorption tolerates temperature fluctuations.
  2. Elements with low-energy excited states (Na, K, Li, Ca) emit strongly even in cool flames, which is why flame photometry is the routine method for Na and K in clinical and soil samples. Characteristic lines: Na 589 nm (yellow), K 766 nm (deep red; the flame itself looks lilac), Li 671 nm (crimson).
  3. Elements whose resonance lines lie deep in the UV (high ΔE, such as Zn) emit poorly in flames; AAS or a hotter source is better.

Inductively coupled plasma (ICP)

An argon plasma sustained by a radio-frequency coil reaches temperatures several thousand kelvin higher than the air–acetylene and nitrous oxide–acetylene flames used in AAS. Consequences:

  • almost complete atomisation; chemical interferences (Ca–phosphate) are largely removed;
  • excited populations are large, so many elements, including refractory ones, emit strongly;
  • the high electron density from ionised argon suppresses ionisation interference of the analyte;
  • multi-element analysis: one spectrum, many lines, measured simultaneously;
  • a wide linear range (several orders of magnitude);
  • the plasma ions can also be fed to a mass spectrometer (ICP-MS), giving very low detection limits and isotope ratios.

The price: spectral interferences multiply because so many lines are emitted, and the argon consumption is high.

AAS versus emission at a glance

FeatureAASFlame emissionICP emission
Measuresground-state atoms (absorption)excited atoms (emission)excited atoms and ions
Temperature sensitivitylowhighlow (stable plasma)
Source neededelement-specific lampnonenone
Elements per runone (per lamp)one or a fewmany simultaneously
Best formost metals at ppm (flame) or ppb (furnace)Na, K, Li, Camulti-element surveys, refractory elements

5. Molecular Fluorescence

After absorbing, a molecule relaxes to the lowest vibrational level of S₁ and may then emit (fluorescence, S₁ → S₀, nanoseconds) or cross to T₁ and emit more slowly (phosphorescence, T₁ → S₀, milliseconds to seconds). Emission occurs at longer wavelength than absorption (Stokes shift).

Intensity and concentration

$$F = k\,\phi\,P_0\left(1 - 10^{-\varepsilon b c}\right) \approx 2.303\,k\,\phi\,P_0\,\varepsilon\,b\,c \quad (\text{when } A < 0.05)$$

φ is the fluorescence quantum yield. Points that follow:

  • Fluorescence is linear in c only at low absorbance. At higher c the excitation beam is absorbed before it reaches the viewed region (inner-filter effect) and the plot bends over.
  • Signal is proportional to \( P_0 \): a stronger lamp gives more signal. This, and measuring against a dark background rather than a small difference between two large signals, is why fluorimetry detects concentrations 10 to 1000 times lower than absorption spectrophotometry.
  • Fluorescence is favoured by rigid, planar, conjugated structures (fluorescein, quinine) and by electron-donating substituents; heavy atoms (Br, I) and paramagnetic species (O₂) promote intersystem crossing and quench fluorescence. Inorganic analytes are measured through fluorescent chelates (Al³⁺ with morin or 8-hydroxyquinoline-type ligands).

Quenching: the Stern–Volmer equation

For dynamic (collisional) quenching by Q:

$$\boxed{\frac{F_0}{F} = 1 + K_{SV}[\text{Q}] = 1 + k_q\,\tau_0\,[\text{Q}]}$$

\( \tau_0 \) is the lifetime without quencher and \( k_q \) the bimolecular quenching constant. A plot of \( F_0/F \) against [Q] is a straight line of slope \( K_{SV} \).

TestDynamic (collisional)Static (ground-state complex)
Lifetimedecreases with [Q]: \( \tau_0/\tau = F_0/F \)unchanged
Temperature risemore quenching (faster diffusion)less quenching (complex dissociates)
Absorption spectrum of fluorophoreunchangedmay change

A fitted \( k_q \) far larger than the diffusion limit (of the order of 10¹⁰ L mol⁻¹ s⁻¹ in water) shows that collisional quenching alone cannot explain the data: some static quenching or binding is present.

6. Other Optical Methods in One Table

MethodMeasuresUse
Turbidimetrydecrease in transmitted light due to scattering by suspended particlessulfate as BaSO₄ suspension
Nephelometryscattered light at 90°very dilute suspensions
Chemiluminescencelight from a chemical reaction, no lampluminol (with H₂O₂ and a catalyst), NO with O₃
X-ray fluorescencecharacteristic X-rays after inner-shell ionisationnon-destructive elemental analysis of solids

Structural spectroscopies (IR, NMR, EPR, Mössbauer, mass spectrometry) are separate units of the Chemical Sciences syllabus; here we treat spectroscopy as a tool for how much, not what structure.

Memory Tricks

Don’t Confuse

ThisvsThat
Absorbance A = −log T (no units)transmittance T (fraction) or %T
Molar absorption coefficient ε (L mol⁻¹ cm⁻¹)specific absorptivity (L g⁻¹ cm⁻¹, mass-based)
Chemical deviation (equilibrium shift on dilution)instrumental deviation (stray light, polychromatic light)
Releasing agent (binds the interferent)protective agent (binds the analyte)
Ionisation suppressor (K, Cs added in excess)releasing agent (La, Sr)
Standard addition: corrects matrix effects on slopeinternal standard: corrects instrument fluctuations by signal ratio
AAS: element lamp, ground-state atomsflame emission: no lamp, excited atoms
Dynamic quenching: τ fallsstatic quenching: τ unchanged
Fluorescence (S₁ → S₀, ns)phosphorescence (T₁ → S₀, ms to s)

Common Mistakes

Practice

All questions below are original practice questions written for this page. They are not from any CSIR NET paper.

Level 1: Foundation

  1. What absorbance corresponds to 1% transmittance?
  2. Name the light source used in AAS and say why a continuum source is unsuitable on its own.
  3. In a Job’s plot, the maximum lies at mole fraction of ligand 0.75. What is the stoichiometry?
Answers
  1. A = −log(0.01) = 2.00.
  2. A hollow-cathode lamp of the analyte element. Atomic absorption lines are far narrower than the bandpass of a monochromator, so a continuum source would be absorbed only in a tiny fraction of the band and the absorbance would be very small and non-linear.
  3. ML₃, since \( n/(n+1) = 0.75 \) gives n = 3.

Level 2: Part B Style

Practice Q1. A solution with \( \varepsilon = 1.2 \times 10^4 \) L mol⁻¹ cm⁻¹ shows A = 0.45 in a 1 cm cell. Its concentration and percentage transmittance are:

(a) 3.75 × 10⁻⁵ M, 35.5% (b) 3.75 × 10⁻⁵ M, 64.5% (c) 5.4 × 10⁻³ M, 35.5% (d) 1.63 × 10⁻⁵ M, 35.5%

Answer: (a)

\( c = 0.45/(1.2 \times 10^4 \times 1) = 3.75 \times 10^{-5} \) M. \( T = 10^{-0.45} = 0.355 \). Option (b) gives the fraction absorbed; (d) uses ln instead of log (0.45/2.303 ÷ 1.2 × 10⁴ ≈ 1.6 × 10⁻⁵ M).

Practice Q2. In the AAS determination of calcium in a phosphate-rich sample, adding LaCl₃ to all solutions:

(a) suppresses ionisation of calcium (b) binds phosphate preferentially and releases calcium for atomisation (c) corrects for background absorption (d) increases flame temperature

Answer: (b)

La³⁺ is a releasing agent: it forms a more stable phosphate than Ca²⁺ does, freeing calcium to atomise. Ionisation suppression (a) needs K or Cs; background correction (c) needs a D₂ lamp or Zeeman system.

Practice Q3. Readings for a single standard addition: 10.0 mL sample diluted to 50.0 mL, A = 0.240; 10.0 mL sample + 5.00 mL of 10.0 ppm standard diluted to 50.0 mL, A = 0.400. The analyte concentration in the original sample is:

(a) 1.50 ppm (b) 5.00 ppm (c) 7.50 ppm (d) 15.0 ppm

Answer: (c) 7.50 ppm

\( 0.400/0.240 = (10.0c_x + 50.0)/(10.0c_x) \Rightarrow 16.67c_x = 10.0c_x + 50.0 \Rightarrow c_x = 7.50 \) ppm. Option (a) is the concentration in the 50 mL flask (7.50 × 10/50).

Practice Q4. Which statement about flame atomic emission compared with AAS is correct?

(a) Emission is less sensitive to flame temperature because it measures ground-state atoms. (b) Emission needs a separate lamp for each element. (c) A small rise in flame temperature changes the emission signal far more than the absorption signal. (d) Alkali metals emit weakly because their excited states lie high.

Answer: (c)

The excited population depends exponentially on T while the ground-state population is almost all the atoms. (a) and (b) describe AAS backwards; (d) is false because alkali-metal resonance lines lie in the visible and are easily excited.

Level 3: Part C Style

Practice Q5. A mixture of X and Y in a 1.00 cm cell gives A = 0.530 at λ₁ and A = 0.420 at λ₂. Molar absorption coefficients (L mol⁻¹ cm⁻¹): X, 8000 at λ₁ and 2000 at λ₂; Y, 1000 at λ₁ and 6000 at λ₂. The concentrations are:

(a) [X] = 6.0 × 10⁻⁵ M, [Y] = 5.0 × 10⁻⁵ M (b) [X] = 5.0 × 10⁻⁵ M, [Y] = 6.0 × 10⁻⁵ M (c) [X] = 6.6 × 10⁻⁵ M, [Y] = 7.0 × 10⁻⁵ M (d) [X] = 6.0 × 10⁻⁵ M, [Y] = 7.0 × 10⁻⁵ M

Answer: (a)
$$8000x + 1000y = 0.530, \qquad 2000x + 6000y = 0.420$$

Multiply the first by 6: \( 48000x + 6000y = 3.180 \). Subtract the second: \( 46000x = 2.760 \), so \( x = 6.0 \times 10^{-5} \) M. Then \( 1000y = 0.530 - 0.480 = 0.050 \), \( y = 5.0 \times 10^{-5} \) M.

Check: \( 2000(6.0 \times 10^{-5}) + 6000(5.0 \times 10^{-5}) = 0.120 + 0.300 = 0.420 \). Option (c) treats each wavelength as if only one component absorbed there (0.530/8000 and 0.420/6000), ignoring the overlap.

Practice Q6. For the sodium 589 nm line (\( g_j/g_0 = 3 \)), consider the statements:

  1. At 2500 K the ratio \( N_j/N_0 \) is about 1.7 × 10⁻⁴.
  2. Raising the flame from 2500 K to 2600 K raises the emission intensity by roughly 45%.
  3. The same rise changes the AAS absorbance by roughly 45%.
  4. An ICP source would give a larger excited-state fraction than the flame.

Which are correct? (Take h = 6.626 × 10⁻³⁴ J s, c = 2.998 × 10⁸ m s⁻¹, k = 1.381 × 10⁻²³ J K⁻¹.)

(a) 1, 2 and 4 only (b) 1 and 3 only (c) 2 and 4 only (d) 1, 2, 3 and 4

Answer: (a) 1, 2 and 4 only

ΔE = hc/λ = 3.37 × 10⁻¹⁹ J.

  • 2500 K: ΔE/kT = 9.77; \( 3e^{-9.77} = 1.71 \times 10^{-4} \). Statement 1 true.
  • 2600 K: ΔE/kT = 9.39; \( 3e^{-9.39} = 2.50 \times 10^{-4} \). Ratio 2.50/1.71 = 1.46. Statement 2 true.
  • N₀ changes from 0.99983 to 0.99975 of the atoms: the absorbance is essentially unchanged. Statement 3 false. (Atomisation efficiency could still change, but not through the Boltzmann factor.)
  • The plasma is much hotter, so the exponential factor is much larger. Statement 4 true.

Practice Q7. The fluorescence of a dye (lifetime τ₀ = 5.0 ns in the absence of quencher) falls to F₀/F = 2.5 when 0.010 M of an iodide salt is added. The Stern–Volmer plot is linear. Which conclusions follow?

  1. K_SV = 150 L mol⁻¹.
  2. If quenching were purely dynamic, k_q would be 3.0 × 10¹⁰ L mol⁻¹ s⁻¹.
  3. This k_q exceeds the typical diffusion limit in water, suggesting a static (binding) contribution.
  4. Purely static quenching would reduce the measured lifetime to 2.0 ns.

(a) 1, 2 and 3 only (b) 1 and 4 only (c) 2, 3 and 4 only (d) 1, 2, 3 and 4

Answer: (a) 1, 2 and 3 only
  • \( K_{SV} = (2.5 - 1)/0.010 = 150 \) L mol⁻¹. True.
  • \( k_q = K_{SV}/\tau_0 = 150/(5.0 \times 10^{-9}) = 3.0 \times 10^{10} \) L mol⁻¹ s⁻¹. True.
  • This is above the order-of-magnitude diffusion limit of about 10¹⁰ in water, so collisions alone cannot account for all the quenching. True (a lifetime measurement would settle it).
  • Static quenching leaves the lifetime of the uncomplexed fluorophore unchanged at 5.0 ns. A lifetime of 2.0 ns (= 5.0/2.5) would be the purely dynamic case. False.

Quick Revision Table

IdeaFormula or fact
Beer–LambertA = −log T = εbc; additive over components
ConversionsT 10% → A 1; T 1% → A 2; T 50% → A 0.301
Best precisionA = 0.434 (T = 36.8%) for constant ΔT; work at 0.2–0.8
Deviationsreal (high c), chemical (equilibria), instrumental (polychromatic, stray light)
Stray lightA’ = −log[(T + s)/(1 + s)]; true 2.00 with 1% stray reads 1.70
LOD3s_blank/slope; LOQ uses 10s
Standard additionspike the sample; corrects matrix (slope) effects
Job’s methodmax at x_L = n/(n + 1)
AAS sourcehollow-cathode lamp of the analyte
AAS interferenceschemical: La/Sr releasing, EDTA protecting; ionisation: K/Cs suppressor; background: D₂, Zeeman
BoltzmannN_j/N_0 = (g_j/g_0) e^(−ΔE/kT); Na 589 nm at 2500 K ≈ 1.7 × 10⁻⁴
Flame photometryNa 589, K 766, Li 671 nm
ICPAr plasma, very hot; multi-element, few chemical interferences, many spectral lines
FluorescenceF ∝ φP₀εbc at A < 0.05; inner-filter at high c
Stern–VolmerF₀/F = 1 + K_SV[Q]; K_SV = k_q τ₀; dynamic lowers τ, static does not

Key Takeaways